Journal of Applied Sciences ›› 1993, Vol. 11 ›› Issue (3): 265-270.
• Articles • Previous Articles Next Articles
SHI TAIHUA, XU SHUJUN, LU YUQING
Received:
Revised:
Online:
Published:
Abstract: The magnetic device is the key part of the nuclear magnetic resonance imaging instrument (NMR-CT). We designed this device by computer aided design and then constructed the imitative device. Experiments show that the difference between the calculation and measurement is only 5.1%, and the field homogeneity is about 2088×10-6.
Key words: nuclear magnetic resonance imaging, magnetic torque, field homogeneity, permanent magnet
SHI TAIHUA, XU SHUJUN, LU YUQING. DESIGN OF THE MAGNETIC DEVICE FOR NUCLEAR MAGNETIC RESONANCE IMAGING INSTRUMENT USED IN MEDICINE[J]. Journal of Applied Sciences, 1993, 11(3): 265-270.
0 / / Recommend
Add to citation manager EndNote|Reference Manager|ProCite|BibTeX|RefWorks
URL: https://www.jas.shu.edu.cn/EN/
https://www.jas.shu.edu.cn/EN/Y1993/V11/I3/265