Diameter Control Technology of Ultra-fine Sutures Based on DMA and Improved GPC Algorithm

WANG Deming1,2, SUI Xiuwu1,2, ZHANG Yang1,2, WAN Kaixin1,2, SHI Feng1,2   

  1. 1. School of Mechanical Engineering, Tianjin Polytechnic University, Tianjin 300387, China;
    2. Tianjin Key Laboratory of Advanced Mechatronics Equipment Technology, Tianjin Polytechnic University, Tianjin 300387, China
  • Received:2018-05-22 Revised:2018-09-29 Online:2019-05-31 Published:2019-05-31

Abstract: In order to solve the problem of suture diameter inhomogeneity caused by slow and imprecise wire diameter detection and uncontrollable forming process, a control system is designed by a rapid measurement and accurate control method. The system, based on the embedded system STM32, consists of a linearly coupled charge-coupled device (CCD) image sensor, 14-bit CCD signal processor AD9822, direct memory access (DMA), and improved adaptive generalized predictive control (GPC) algorithm. The suture samples satisfying the requirements of the 32th United States Pharmacopoeia Suture standard have been produced on the JK1601 spinning machine by the control system, and the uniformity of the wire diameter has increased from 74.6% to 95%.

Key words: charge-coupled device (CCD), embedded system, ultra-fine medical absorbable sutures, AD9822, improved generalized predictive control (GPC) algorithm